This kit includes the following below (serial number may not be exact as shown in photos).
QTY: 1, 50 ohm EFT attenuator (Input Connector: SHV)
QTY: 1, 1000 ohm EFT attenuator (Input Connector: SHV)
QTY: 1, SHV Plug (Male) to BNC Male Adapter
QTY: 1, 50 ohm RG58 Cable (Length: 5ft)
QTY: 1, Electro Calibration Lab carrying case for EFT Attenuator set
UPC/EAN: 860012796212
Meets EN/IEC 61000-4-4 & ISO7637
ECL50 (F) is designed for pulse calibration with a 50Ω load at the front input and a matching 50Ω output impedance which meets and exceeds EN/IEC 61000-4-4 & ISO7637. While the specification exceeds 400MHz (875ps rise time), our design at the 3dB bandwidth is over 500MHz achieving 700ps to ensure you have no doubt of capturing your pulse rise time (10% - 90%). This item mates directly to generators utilizing SHV connector.
Specification
Max. power rating: 5 Wrms
Max. peak voltage (burst/EFT pulses as specified in EN/IEC 61000-4-4): 10.0 kV
Max. operating time at 10.0 kV: 300 s (5 kHz, 300ms/15ms)
Nominal attenuation (terminated by 50 Ω): -60dB
Nominal voltage ratio (See exact values on labels): 1000:1
Input Impedance: 50 Ω ±2%
Output Impedance: 50 Ω ±1%
Bandwidth (3 dB): >400 MHz
Connector Input: SHV connector, socket
Connector Output: BNC jack, female
UPC/EAN: 860012796205
- Meets EN/IEC 61000-4-4 & ISO7637
ECL1000 is designed for pulse calibration with a 1000Ω load at the front input and a matching 50Ω output impedance which meets and exceeds EN/IEC 61000-4-4 & ISO7637. While the specification exceeds 400MHz (875ps rise time), our design at the 3dB bandwidth is over 800MHz achieving 437.5ps to ensure you have no doubt of capturing your pulse rise time (10% - 90%). This item mates directly to generators utilizing Fischer F103 connector.
Specification
Max. power rating: 5 Wrms
Max. peak voltage (burst/EFT pulses as specified in EN/IEC 61000-4-4): 10.0 kV
Max. operating time at 10.0 kV: 300 s (5 kHz, 300ms/15ms)
Nominal attenuation (terminated by 50 Ω): -66dB
Nominal voltage ratio (See exact values on labels): 2000:1
Input Impedance: 1000 Ω ± 2% // ≤ 6 pF
Output Impedance: 50 Ω ±1%
Bandwidth (3 dB): >400 MHz
Connector Input: SHV connector, socket
Connector Output: BNC jack, female